Attachement 1 to the Agreement - Order description
|
Attachement 1 to the Agreement - Order description
Components of 32-channel time correlated single photon counting detecting system
|
|
||||
|
Technical parameter |
The required value of the technical parameter |
The value of the technical parameter offered by the Supplier and confirmation of the functional feature of the system component offered by the Supplier
|
||
1. DETECTION SETUP |
|
||||
1.1 |
Number of detection channel |
32 |
|
||
1.2 |
Number of TCSPC units |
32 |
|
||
1.3 |
Maximum number of TCSPC MODULES |
4 |
|
||
1.4 |
Connection of TCSPC MODULES with PC unit |
USB lub PCIExpress |
|
||
1.5 |
TCSPC MODULES mounting |
19’’ RACK |
|
||
1.6 |
Time resolution (electronic) |
< 12 ps |
|
||
1.7 |
Minimum acquisition time of single DTOF |
< 25 ms |
|
||
1.8 |
Maximum countrate |
> 4 MHz |
|
||
1.9 |
Number of single photon detectors |
32 |
|
||
1.10 |
32 detectors power supply ( MODUŁ POWER) |
Dedicated 32-channel power supply or 2 dedicated 16-channel power supplies |
|
||
1.11 |
Spectral range of the detectors (minimum range) |
600-900nm |
|
||
1.12 |
Stability (shift of the maximum of instrumental response function (IRF) during the time of measurement) |
< 10ps |
|
||
1.13 |
Stabilization time of the detecting system |
< 30 minutes |
|
||
1.14 |
Diameter of an active detector area |
≥ 7 mm |
|
||
1.15 |
Full width at half maximum of the IRF (with full area of the cathode) |
< 180 ps |
|
||
1.16 |
Overload protection of the detectors |
YES |
|
||
1.17 |
Reset of overload protection of the detectors |
SOFTWARE operated, functionality of resetting all detectors simultaneously must be implemented |
|
||
2. Laser light source EMITTING SETUP |
|
||||
2.1 |
Repetition rate |
>= 50MHz |
|
||
2.2 |
Number of pulsed laser light sources @ 1 |
4 |
|
||
2.3 |
Number of pulsed laser light sources @ 2 |
4 |
|
||
2.4 |
Wavelength 1 |
From range 670 nm-730 nm |
|
||
2.5 |
Wavelength 2 |
From range 830 nm-850 nm |
|
||
2.6 |
Optical power |
> 10 mW |
|
||
2.7 |
Fiber coupler |
YES (SMA or FC/APC @ fiber core400 m NA=0.2 |
|
4