Analytical Balance. The analytical balance used to determine the weights of all filters shall have a precision (standard deviation) of 2 μg and a resolution of 1 μg (1 digit = 1 μg) specified by the balance manufacturer.
Appears in 2 contracts
Samples: Agreement Concerning the Adoption of Uniform Technical Prescriptions for Wheeled Vehicles, Agreement Concerning the Adoption of Uniform Technical Prescriptions for Wheeled Vehicles
Analytical Balance. The analytical balance used to determine the weights of all filters shall have a precision (standard deviation) of 2 μg 20µg and a resolution of 1 μg 10µg (1 digit = 1 μg) specified by 10µg). For filters less than 70 mm diameter, the balance manufacturerprecision and resolution shall be 2µg and 1µg, respectively.
Appears in 1 contract
Analytical Balance. The analytical balance used to determine the weights of all filters shall have a precision (standard deviation) of 2 μg 20μg and a resolution of 1 μg 10μg (1 digit = 1 μg) specified by 10μg). For filters less than 70 mm diameter, the balance manufacturerprecision and resolution shall be 2μg and 1μg, respectively.