Exhibit 10.6
Portal Manufacturing Agreement
By and between
Diametrics Medical, Inc.
0000 Xxxxxx Xxxx
Xx. Xxxx, XX 00000, XXX
(hereinafter referred to as "Diametrics")
and
PHILIPS Medical Systems
(hereinafter referred to as "Philips")
on the supply of Portal cartridge reader M3561A (M3561-60001);
exchange repair unit M3561-68000
plus accessories edge connector cleaning kit M3672A (DMI 450000);
temp card M3565A (DMI 467900); edge connector M3561-66403
(DMI 516300); and DeviceSet with DeviceCom on a CD M3564A (DMI464218)
Portal Manufacturing Agreement
THIS PORTAL MANUFACTURING AGREEMENT (the "Manufacturing Agreement"), made
as of this 10th day of April, 2003 (the "Effective Date"), by and among
Diametrics Medical Incorporated, a Minnesota corporation having its principal
place of business at 0000 Xxxxxx Xxxx, Xx. Xxxx, Xxxxxxxxx 00000 ("Diametrics")
and Philips Medical Systems North America Company, a division of Philips
Electronics North America Corporation, a Delaware corporation with offices at
0000 Xxxxxxxxx Xxxx, Xxxxxxx, Xxxxxxxxxxxxx ("Philips"). Diametrics and Philips
shall each be a "Party."
RECITALS
A. Philips and Diametrics previously entered into a distribution
agreement dated June 6, 1999 (the "Distribution Agreement") the term of which
expired on October 31, 2002, and the parties are entering into an amendment to
the Distribution Agreement, dated as of the date of this Manufacturing Agreement
(the "Amendment") in order to clarify some of the surviving rights and
obligations under the Distribution Agreement. All capitalized terms shall have
the meanings set forth in this Manufacturing Agreement.
B. Whereas as governed by the Distribution Agreement, Philips and
Diametrics have jointly developed the M3650A Blood Analysis Portal System
consisting of the M3561A Portal Reader and the M3562A Interface Module.
C. Under the terms of the Distribution Agreement, Philips has surviving
rights to distribute the Portal Measurement Module as a part of the Portal
Product in a specific field of use until October 31, 2004, and Diametrics will
manufacture the Portal Measurement Products in accordance with the Amendment and
this Manufacturing Agreement and provide Portal Measurement Products to Philips
to allow for the completing of the postponement step by Philips of the Portal
Products.
C. The Parties desire to enter into this Manufacturing Agreement for the
purposes of Diametrics' manufacture of the Portal Measurement Products solely
for Philips on the terms of this Manufacturing Agreement.
NOW THEREFORE, in consideration of the foregoing premises and of the mutual
covenants and subject to the terms and conditions set forth herein below,
Philips and Diametrics agree as follows:
1. Definitions
1.1 DeviceSet/DeviceCom Software. Means DeviceSet software with DeviceCom
software on a CD M3564A (DMI 464218), as more particularly described
in the Specifications.
1.2 Portal Measurement Accessories. Means the following (a) accessories
edge connector cleaning kit M3672A (DMI 450000); (b) temp card M3565A
(DMI 467900); (c) edge connector M3561-66403 (DMI 516300), and (d)
M3561-40003
1
edge connector cover (DMI 466400), as more particularly described in
the Specifications.
1.3 Portal Measurement Module. Means the Portal cartridge reader M3561A
(M3561-60001), as more particularly described in the Specifications.
1.4 Portal Measurement Products. Means the Portal Measurement Module, the
exchange repair unit M3561-68000, Portal Measurement Accessories and
DeviceSet/DeviceCom Software, collectively.
1.5 Portal Products. Has the meaning set forth in Section 1.14 of the
Amendment.
1.6 Required Quality Data. Means the data and information described on
Schedule 1.
1.7 Specifications. Means the specifications set forth in Schedule 2.
1.8 Term. Means the period from the Effective Date through October 31,
2004.
1.9 Schedule References. All Schedule references shall mean the Schedules
to this Amendment.
2. Effect of Agreement
2.1 Portal Manufacturing Agreement. This Manufacturing Agreement
stipulates the additional terms and conditions for the supply of
Portal Measurement Products to Philips by Diametrics, including any
Portal Measurement Products supplied by Diametrics prior to or during
the Term hereof. The Portal Measurement Products will be manufactured
and supported by Diametrics as provided herein and the Amendment.
2.2 Supplementary Agreement. The Parties are executing this Manufacturing
Agreement as a supplement to the Amendment in order to provide for
certain additional terms and conditions relating to the manufacture
of the Portal Measurement Products. This Manufacturing Agreement does
not supercede the Amendment, and the terms of the Amendment shall
apply to the Portal Measurement Products, except as expressly set
forth herein.
3. Examination and delivery of the Diametrics production units prior to
shipment to Philips.
3.1 Supplier Evaluation. Diametrics shall manufacture the Portal
Measurement Products with material purchased directly by Diametrics.
A list of suppliers and contacts have been provided by Philips to
Diametrics. Philips agrees to provide initially the "Supplier
Evaluation". Diametrics' and Philips' engineering contacts (listed on
Schedule 5) will jointly agree as to which supplied components of the
Portal Measurement Module are required to be periodically audited by
the Diametrics quality personnel.
2
3.2 Certificate of Compliance. Each serialized Portal Measurement Module
will be provided with a completed Diametrics "Certificate of
Compliance" certifying that said product was inspected and tested per
controlled manufacturing procedures.
3.3 Production Units. Diametrics shall produce units of Portal
Measurement Modules and other Portal Measurement Products in
accordance with the terms and provisions hereof. The scope of
delivery and supply shall be as set forth in this Manufacturing
Agreement and the Amendment.
3.4 Audit by Philips. Philips may periodically audit via a functional
test a sample of the delivered Portal Measurement Module. Should
Philips detect deviations from the Specifications or defects during
the course of the functional check of the full production Portal
Measurement Modules, Philips shall inform Diametrics thereof
forthwith in writing via an SAR ( Supplier Action Request) per the
Supplier Action Request Process Q2920-00224. In this event,
Diametrics shall provide Philips with repaired or newly manufactured
units of Portal Measurement Modules which are free of the defect(s)
in question and shall provide a written corrective action plan within
a reasonable period of time to be defined and agreed to by the
Parties. Diametrics shall ensure that defects are remedied and
rectified by certified employees. Subject to mutual agreement between
the Parties, such rectification of defects may be carried out by
Philips.
3.5 Functional Checks after Delivery. Following delivery of the repaired
/ newly manufactured units, Philips shall conduct another functional
check in accordance with Section 3.4 above. If deviations from the
Specifications or defects are still detected in the repaired/newly
manufactured Portal Measurement Modules and these are not remedied
and rectified within mutually agreed upon date negotiated within the
SAR process, subject to mutual agreement of the Parties, such
rectification of defects may be carried out by Philips or a third
party named by Diametrics and agreed by Philips prior to the
implementation at Diametrics expense.
3.6 Ship to Stock. The quality of the Portal Measurement Modules will
allow ship-to-stock, without need of incoming inspection/test.
Philips may periodically audit via a functional test a sample of the
delivered Portal Measurement Module with a goal of acceptable failure
rate of 1 per 100 units.
4. Scope of Delivery.
4.1 Forecasting and Ordering. Philips shall issue to Diametrics a monthly
written, rolling, non-binding six (6) month forecast of its purchase
of Portal Measurement Products. Such forecasts shall be updated
monthly. Philips shall issue binding written purchase orders for
Portal Measurement Products at least sixty (60) days prior to the
proposed shipping date for such items. All purchase orders in excess
of Philips' forecast shall be subject to written acceptance by
Diametrics. Diametrics will use reasonable efforts to supply such
items in a timely fashion,
3
subject to Diametrics' reasonable production capabilities and
provided that such purchase orders are consistent with Philips'
forecasts.
4.2 Fluctuations in Orders. The parties shall address purchasing of
components and materials as provided in Section 4.1(c) of the
Amendment. In addition, Diametrics will diligently monitor its supply
chain and its production processes and will promptly inform Philips
of any anticipated interruption or inability to supply Portal
Measurement Products to Philips to meet its forecasts or its firm,
binding purchase orders, in order to permit Philips to assess the
manufacturing situation, to inform its sales and planning personnel
to attempt to adjust for such shortfall, and to advise and assist
Diametrics, at Diametrics' expense in its efforts to minimize or
avert such interruption in or inability to supply Portal Measurement
Products.
5. Delivery.
5.1 Purchase Orders for Portal Measurement Products. Philips shall be
entitled to use its then current standard purchase order for its
purchases hereunder. In the event of a conflict between the terms of
any such purchase order and the terms of this Manufacturing
Agreement, the terms of this Manufacturing Agreement shall take
precedence. Additional terms included in the purchase order that are
not terms of this Manufacturing Agreement will be subject to prior
mutual written agreement.
5.2 Terms for Delivery. All deliveries of Portal Measurement Products and
Consumables shall be made by Diametrics F.O.B. Diametrics U.S. or
European facility. Title and risk of loss shall pass to Philips at
the time of tender at Diametrics' facility to the carrier designated
by Philips. Upon delivery to Philips' designated carrier, Philips
will assume title and risk of loss, and will be responsible for
transportation, and, if applicable, export of such products from the
country of manufacture and shall be entitled to any duty drawback for
which the products qualify. Diametrics shall provide Philips, upon
Philips' reasonable request and at Philips' expense, reasonable
substantiation and assistance with respect to such duty drawbacks.
The pricing for products excludes, and Philips shall pay, all
shipping, handling and insurance costs for and other costs of
transporting such products after delivery to the F.O.B. point.
5.3 Diametrics Supply Obligation. Diametrics shall manufacture the Portal
Measurement Products in accordance with the applicable specifications
and in accordance with the FDA, EU and all applicable regulations.
6. Prices.
6.1 Prices. Portal Measurement Products shall be supplied at the prices
set forth in Schedule 4. All prices are F.O.B. Diametrics
manufacturing facility.
6.2 Price Changes. Prices may be changed upon ninety (90) days written
notice by Diametrics. Any changes will not impact orders received
prior to the date of such change.
4
6.3 Terms of Payment. All payments for Portal Measurement Products shall
be due and payable within thirty-five (35) days of the date of the
applicable invoice. Philips shall make all payments in immediately
available funds to the location and in the manner reasonably
designated by Diametrics from time to time. A late fee shall be paid
by Philips on any amount not received by Diametrics when due at a
rate of 1.5% per month on all unpaid amounts, or the maximum rate
permitted by law, whichever is less. As between the Parties, Philips
shall be responsible for all taxes relating to the Portal Measurement
Products so purchased (except for taxes on Diametrics' net income).
7. Invoice Address.
All of Diametrics' invoices are the addressed to:
Philips Medizin Systeme Boeblingen Gmbh
Financial Service Center
Xxxxxxxx 0000
00000 Xxxxxxxxxx/ Xxxxxxx
8. Contact Persons. Diametrics and Philips shall each designate a Contact
Person who will meet (by phone or in person) as required to discuss and
monitor any manufacturing and manufacturing capacity issues for the Portal
Measurement Products. Both Parties shall appoint additional contact persons
in Schedule 5 to facilitate communication regarding specific aspects of the
manufacture and supply of the Portal Measurement Products hereunder.
9. Term and Termination of this Manufacturing Agreement.
9.1 Term. This Manufacturing Agreement shall become effective on the
Effective Date and shall terminate on October 31, 2004, unless
terminated as provided herein.
9.2 Termination for Material Breach. Either party may terminate this
Manufacturing Agreement if the other party materially breaches this
Manufacturing Agreement and fails to remedy such breach within ninety
(90) days written notice from the non-breaching party. Material
breaches shall include failure by Diametrics to meet the agreed
quality standards for the Portal Measurement Products as set forth in
Schedule 2.
9.3 Effect of Termination. No termination or expiration of this
Manufacturing Agreement shall affect or discharge any obligations,
rights, disclaimers, conditions or limitations of either Party which
arose prior to the effective date of such termination. In addition,
Articles 15 (Confidentiality) and 21 (Miscellaneous) and any Sections
necessary to give effect to this Article 9 shall survive any
termination or expiration of this Manufacturing Agreement.
5
10. Technical Alteration. Diametrics agrees and undertakes to effect technical
alterations (i.e., any required technical changes due to third party
supplier changes, such as part obsolescence, revision changes and the like)
to the Portal Measurement Modules only after written consent of Philips.
11. Product Change Notification. No process changes, design changes,
geographical relocation of manufacturing processes, or process step
discontinuances affecting the electrical performance (whether specified or
not), the mechanical form or fit, the environmental compatibility or
material chemical characteristics, or the life reliability of Portal
Measurement Products, but excluding any changes that do not affect the fit,
form or function of the Portal Measurement Products (hereinafter
collectively called "Changes") shall be made or incorporated in Portal
Measurement Products without following the change notification process
established herein. Diametrics shall give Philips written notice of any
proposed change via the established Philips/Diametrics change order process
and provide evaluation samples and other appropriate information as
specified by Philips. For planned/scheduled changes that require
re-qualification by Philips, Diametrics notice must be received by Philips
at least three (3) months prior to the first proposed shipment of any such
Products. For changes not requiring re-qualification by Philips, Diametrics
notice must be received by Philips at least one (1) month prior to the
first proposed shipment unless mutually agreed by the Parties. The Supplier
Change Notification Agreement process Q2920-00094 will be followed. All
product changes will be reflected in the 10 digit serial number by an
appropriate change of the year (digit 3) and/or week of the change (digit
4-5) as appropriate.
12. Replacement Requirements/Exchange Products. Diametrics guarantees to
Philips that replacement and or exchange Portal Measurement Modules will be
supplied for a further seven (7) years following delivery of the last
series/batch. After expiration of this seven (7) year period, Diametrics
shall make available replacements (backward compatible or equivalent) in
accordance with the respective technical possibilities or give Philips the
possibility of placing a final order of a reasonable quantity, thus
guaranteeing Philips a supply up until the end of the service period. The
prices, terms and conditions as well as the technical feasibility of these
two cases will form the subject of a separate agreement at the time of
delivery of the last series/batch.
13. Repaired / Exchanged Products.
Diametrics will provide a test and repair service for Portal Measurement
Modules returned from the customers (internal or external) as faulty
whether still in warranty or out of warranty. Philips will provide
Diametrics' field failure information. Philips will pay Diametrics the
exchange repair cost per Schedule 4. Diametrics shall perform root cause
analysis (sub-assembly level) on any failed unit in order to determine
failure mode. Diametrics will provide a pareto trend analysis of failure
data for all failed units to Philips on a monthly basis.
14. Documentation
6
14.1 Rights of Access. Philips shall have the right to access and use
Diametrics' manufacturing documentation for the Portal Measurement
Products to comply with regulatory requirements.
14.2 Record Retention. Diametrics shall retain records of any safety,
quality and reliability related data relating to the Portal
Measurement Products as well as any data necessary for "Current Good
Manufacturing Practices" (CGMP, as required by FDA), the "In Vitro
Diagnostics Directive" (the IVDD, as required by the European Union)
and the "Medical Device Directive" (MDD, as required by the European
Union), to the extent provided under Schedule 1, Required Quality
Data.
14.3 Access to Filings. Diametrics assures the access of DHR (Device
History Record), DHT (Device History Traveler), and DMR (Device
Master Record) in support of Philip's timely answering of Portal
Measurement Product questions from the FDA. Diametrics shall
cooperate with Philips in connection with its regulatory obligations
to provide to Philips, on a timely basis, the necessary reports
relating to complaints and product performance issues regarding the
Portal Measurement Products. Upon reasonable request, Diametrics
shall promptly make such reports (on behalf of Philips) directly to
the appropriate regulatory authorities.
15. Confidentiality.
15.1 Non-Use and Non-Disclosure. Each Party acknowledges and agrees that
all the other Party's Confidential Information is confidential to the
disclosing Party. Each Party shall take the same reasonable measures
as it uses to protect its own confidential information from the
unauthorized disclosure or misuse to protect the other Party's
Confidential Information from unauthorized disclosure or misuse,
including without limitation, any disclosure by its employees,
agents, contractors, permitted sublicensees, or consultants of the
other Party's Confidential Information. As used herein, the term
reasonable measures shall mean at least those measures a Party
applies to the protection of its own Confidential Information and the
term misuse shall mean use for any purpose other than as permitted or
required hereunder.
15.2 Marking. To be entitled to protection as Confidential Information,
all Diametrics or Philips documents containing that Party's
Confidential Information shall be appropriately and clearly marked as
"Proprietary," "Secret," "Confidential," or other words to similar
effect. If a disclosure of Confidential Information is made orally,
as in a meeting, the disclosing Party shall indicate the nature of
that information at the time of its disclosure and shall confirm such
designation in writing within ten (10) days of the date of such
disclosure to the receiving Party.
15.3 Exclusions. Information shall not be considered Confidential
Information hereunder if it:
7
(a) was already in the possession of the receiving Party prior to
its receipt from the disclosing Party;
(b) is, or becomes, part of the public knowledge or literature
through no fault, act or omission of the receiving Party,
provided, Confidential Information shall not be deemed to have
entered the public domain by reason of its having been filed
with any regulatory authority; provided the disclosing Party has
taken advantage of any procedures available to protect
confidentiality, including FOIA marking and protective orders;
(c) is, or becomes, available to the receiving Party from a source
other than the disclosing Party, which source has rightfully
obtained the same information and has no obligation of
confidentiality to the disclosing Party with respect to it;
(d) is made available on an unrestricted basis by the disclosing
Party to a third party unaffiliated with the disclosing Party;
or
(e) is required to be revealed pursuant to law or requirements of
any securities exchange on which a Party's shares are listed and
traded, provided, however, the receiving Party which is under
any such requirement of law shall give reasonable notice to the
disclosing Party of such requirement and shall cooperate with
the disclosing Party, at the disclosing Party's expense in
reasonable legal efforts to limit or mitigate any such
revelation so as to preserve the proprietary nature of any
Confidential Information contained therein.
15.4 Duration; Surviving Obligation. This Article 15 shall be deemed to
cover and include any non-public information disclosed by a Party to
the other during the course of their negotiations of this
Manufacturing Agreement, whether or not marked or indicated as
provided in Section 15.2. Each Party's obligations of non-use and
non-disclosure of the other Party's Confidential Information shall
apply during the term of this Manufacturing Agreement and shall also
survive for a period of three (3) years after its termination for any
reason.
15.5 Confidentiality of this Agreement. The terms of this Manufacturing
Agreement itself and the Schedules hereto shall be deemed to be
Confidential Information hereunder. In the event that a Party is
required to disclose the content of this Manufacturing Agreement
pursuant SEC requirements or requirements of any securities exchange
on which a Party's share are listed and traded, such Party shall use
its reasonable efforts to obtain confidential treatment of at least
the terms of Schedules 1, 2, 4, 6 and 7.
16. Warranty.
16.1 Warranty Terms. Diametrics warrants to Philips that the Portal
Measurement Products at the time of their delivery by Diametrics to
Philips (a) shall meet the Specifications as set forth on Schedule 2;
(b) shall have been manufactured in
8
accordance with all laws and regulations applicable to their
manufacture in those jurisdictions in which Philips is distributing
the Portal Measurement Products (provided that Philips gives
Diametrics at least ninety (90) days written notice of any
jurisdictions in addition to those in effect as of the Effective Date
of this Amendment); (c) shall be new or newly manufactured; and (d)
shall be of good and merchantable title, free of liens and
encumbrances. Diametrics shall, promptly, at Diametrics' sole option,
replace, repair or make a purchase price (in the amount paid by
Philips to Diametrics) refund for any of such Portal Measurement
Products proved to be non-conforming, provided that written notice
and reasonable documented evidence of each warranty claim and the
fact that the failure occurred during the warranty period is received
by Diametrics within thirty (30) days after the expiration thereof.
Diametrics shall have the sole right to verify such non-conformance.
Such replacement, repair or refund shall be Philips' sole remedy
hereunder. Unless otherwise agreed by the Parties, the warranty
period under this Section 16.1 for (1) the Portal Measurement Modules
shall be eighteen (18) months from the date of shipment to Philips;
(2) any DeviceSet/DeviceCom Software shall be forty-five (45) days
from the date of delivery of such software to Philips' customer; and
(3) any Portal Measurement Accessory shall be ninety (90) days from
the date of shipment to Philips. If requested by Diametrics, Philips
shall return the non-conforming Portal Measurement Products to
Diametrics at the time of submission of the warranty claim therefore.
Philips agrees to provide Diametrics sufficient notice of additional
countries in which it intends to distribute the Portal Measurement
Products to permit Diametrics to meet its obligations under
subsection (b) of this Section 16.1.
16.2 Exclusions. This warranty does not apply to Portal Measurement
Modules or Portal Measurement Accessories that have been repaired,
modified or tampered with by anyone other than Diametrics qualified
technical personnel, nor to Portal Measurement Modules or Portal
Measurement Accessories that have been exposed (by customer
negligence) to adverse conditions through improper installation,
application or maintenance.
16.3 NFF Procedures. Any Portal Measurement Modules returned to Diametrics
with no fault found (NFF) shall undergo twenty four (24) hour
"burn-in" (i.e. 4 cycles of Treadmill test) upon no fault
determination. Following completion of burn-in, the Portal
Measurement Modules shall be subjected to a full functional test.
Provided no fault is found, the Portal Measurement Modules shall be
labeled (i.e. DHR annotated) prior to return to Philips in order to
identify the Portal Measurement Modules, should they every be
returned again. If the same Portal Measurement Modules should be
returned to Diametrics' facility for a second time, and NFF, the
Portal Measurement Modules shall be analyzed and replaced. For Portal
Measurement Module's second NFF, the pc boards will be replaced and
product fully tested before returned to Philips. If same Portal
Measurement Module is returned a third time and NFF, the entire
Portal Measurement Module will be removed from use and replaced at
Diametrics' expense. The foregoing obligations apply only if the
Portal Measurement Module is under warranty.
9
16.4 Warranty Disclaimer. EXCEPT AS EXPRESSLY PROVIDED IN THIS ARTICLE 16
OF THIS AGREEMENT, DIAMETRICS MAKES NO REPRESENTATIONS OR WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WITHOUT LIMITATION, ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
16.5 Customer Warranties. Philips shall be entitled to warrant the Portal
Measurement Products to its customers on terms it deems appropriate;
provided that Philips shall bear any expenses it may incur with
respect to such warranties, to the extent that such warranties exceed
the scope or term of the express warranties made by Diametrics
hereunder.
16.6 Customer Complaints, Warranty and Out-of Warranty Repair. Philips
will be responsible for handling customer complaints regarding the
Portal Measurement Products. Diametrics will provide reasonable
assistance to Philips in resolving such customer complaints.
Diametrics will accept the return of any Portal Measurement Products
that do not conform to the warranties set forth in Section 16.1, and
will repair or replace any such defective Portal Measurement Product
(or refund Philips' purchase price) as provided in Section 16.1.
Philips agrees to maintain an inventory of Portal Measurement
Products to meet its on-going repair obligations; provided that
Diametrics agrees to maintain a reasonable inventory of Portal
Measurement Products to support significant Product warranty issues.
As between Diametrics and Philips, Philips will be responsible for
providing its customers with such replacement Portal Measurement
Products for use while such defective Products are being repaired or
replaced. Diametrics will also provide out-of-warranty repair service
on the Portal Measurement Modules, on the terms set forth on Schedule
4 for a period of not less than seven (7) years after last production
run of the Portal Measurement Module (or such longer period as may be
required by local laws).
16.7 Return Procedures. Reasonable attempts by Philips 1st Level
Application Support (Response Centers) to determine failure mode(s)
of the Portal Measurement Products is assumed. This information will
be transferred to Diametrics prior to any Portal Measurement Products
being returned from the field. This will provide initial information
to assist in the investigation of the failed Portal Measurement
Product. Upon receipt of a failed Portal Measurement Module, the unit
will go through the standard repair process, which includes the known
repair (if applicable) and then 1 cycle of Treadmill testing (roughly
equivalent to 6 hour "burn-in"), and the subsequent quality tests. If
the failure mechanism cannot be identified the unit will undergo the
quality (diagnostic) tests, the repair pursued and then a full retest
of the Portal Measurement Module (including Treadmill and the quality
tests). If the fault cannot be re-created internally, and it passes
all testing requirements the unit will be sent back out as acceptable
(and documented as no fault found ("NFF") on the DHR). Diametrics
will follow the Supplier Corrective Action Report ("SCAR") process or
Distributor Action Request ("DAR") process to notify Philips if there
is are multiple failures of any one component of any of the Philips
engineered
10
components. Diametrics, likewise will also pursue a Corrective Action
Report ("CAR") if there are multiple failures of any one component of
any Diametrics engineered components. Philips will be notified at the
time of the CAR initiation.
16.8 Repair Period. Diametrics agrees to use commercially reasonable
efforts to repair or replace the non-conforming Portal Measurement
Products and return conforming Portal Measurement Products to Philips
within four (4) weeks of receipt of Diametrics. Failed units returned
from the field that are out-of-warranty will be repaired within 10
working days for customer units and within 30 days for Service
Logistics International ("SLI") exchange units.
17. Quality Performance. If the failure rate of the Portal Measurement Module
exceeds specified limits, (e.g. as per Section 3.5) and/or the Philips
field failure rate is higher than the expectation (i.e. for Portal
Measurement Module less than 4% the first year with a decrease to 1% the
second year), Diametrics commits to initiate an SCAR, DAR or CAR driven
corrective action project for quality / reliability improvements on Portal
Measurement Modules (material, processes, subassemblies) effective per the
SCAR or CAR.
The goal for the Portal Measurement Module is not to exceed an annual
failure rate (AFR) of 4% the first year with a decrease to 1% the second
year for the Portal Measurement Module within a one year rolling period
starting when the contract becomes effective:
AFR = # units failed/ #units shipped by Philips (annualized)
Diametrics and Philips will do a monthly failure clearance on a common
basis before entering the numbers into the calculation.
18. Product Recall.
18.1 If either Party believes that a recall of any Portal Measurement
Product manufactured under this Manufacturing Agreement is desirable
or required by law, it shall promptly notify the other Party. The
Parties shall then discuss reasonably and in good faith whether such
recall is appropriate or required and the manner in which any
mutually agreed recall shall be handled. This Section 18 shall not
limit the obligations of either Party under law with respect to
recall of Portal Measurement Products required by law or properly
mandated by governmental authority. Voluntary recalls shall be
conducted by mutual agreement (with such agreement not to be
unreasonably withheld) provided that if mutual agreement is not
reached, either Party may individually conduct the voluntary recall
in question in a manner consistent with its own regulatory guidelines
and criteria. The "Recalling Party" shall bear all costs and expenses
of any voluntary recall of any Portal Measurement Products and shall
reimburse the "Other Party" for the reasonable, out-of-pocket costs
incurred by the Other Party as a result of such recall, including the
replacement cost of any Portal Measurement Products affected thereby,
unless (i) such recall is unjustified and
11
neither requested nor classified as a recall by a governmental agency
or delegee or (ii) the cause or basis of such recall is attributable
to a condition, fact or action that constitutes (A) a breach by the
Other Party of any of its obligations hereunder or (B) negligence or
willful misconduct of the Other Party in which case the Other Party
will be liable for the costs and expenses of such recall, and shall
reimburse the Recalling Party for the reasonable, out-of-pocket costs
incurred by Recalling Party as a result of such recall, including the
replacement cost of any Product affected thereby. The Parties shall
cooperate fully with each other in effecting any recall of the
Products pursuant to this Section 18, including communications with
any customers or to the public.
18.2 If any governmental agency having jurisdiction (including without
limitation the FDA) shall request or order any corrective action with
respect to Portal Measurement Products supplied hereunder, including
any Portal Measurement Product recall, customer notice, restriction,
corrective action or market action or any Product change, Diametrics
shall bear the costs and expenses of such corrective action and shall
reimburse Philips for the reasonable, out-of-pocket costs incurred by
Philips as a result of such corrective action, including the
replacement cost of any Products affected thereby, to the extent that
the cause or basis of such corrective action is attributable to a
condition, fact or action that constitutes (A) a breach by Diametrics
of any of its obligations hereunder or (B) negligence or willful
misconduct of Diametrics. Philips shall bear all other costs and
expenses of such corrective action, and shall reimburse Diametrics
for the reasonable out-of-pocket costs incurred by Diametrics as a
result of such action, including the replacement cost of any Portal
Measurement Product affected thereby.
19. Environmental Considerations.
Diametrics agrees to provide commercially reasonable support to Philips, at
Philips' written request and at Philip's expense, as required to meet
United States and international environmental laws, regulations and
standards applicable to the manufacture of the Portal Measurement Products.
20. Product Discontinuances. Diametrics acknowledges its obligation to
manufacture the Portal Measurement Products during the term of this
Manufacturing Agreement; however, if the Portal Measurement Products will
not be available from Diametrics prior to the conclusion of the term,
Diametrics shall give written notice to Philips. Diametrics notice shall be
given immediately upon Diametrics knowing about the unavailability, but in
any event no less than six (6) months in advance of the last order date of
the Products. In any event of termination of this Manufacturing Agreement
and subject to Diametrics' manufacturing capacity, and in addition to any
other remedies that Philips may have at law or in equity, Philips shall be
entitled to determine its lifetime-buy quantities and to purchase such
quantities in accordance with this Manufacturing Agreement for purchase and
shipment by the end of the Term.
21. Miscellaneous Provisions
12
21.1 Entire Agreement. This Manufacturing Agreement, together with any
Schedules or other Attachments hereto and the Amendment, constitutes
the entire agreement between the parties in relation to this subject
matter and as such, supersedes all prior and contemporaneous
negotiations, agreements, representations, understandings and
commitments with respect thereto and shall take precedence over all
terms, conditions and provisions on any purchase order form, or order
acknowledgment, or order release purporting to address the same
subject matter.
21.2 Modification. This Manufacturing Agreement shall not be released,
discharged, changed or modified in any manner except by a writing
signed by the duly authorized officers or agents of each Party
hereto, which writing shall make specific reference to this
Manufacturing Agreement and shall express the plan or intention to
modify same.
21.3 Severability. If any provision or provisions of this Manufacturing
Agreement shall be determined to be unenforceable, then the Parties
shall in good faith negotiate for a substitute provision addressing
the same subject matter as the unenforceable provision(s) as may then
be considered to be enforceable, provided, however, if no substitute
provision can be formulated which shall be accepted by the Parties as
enforceable, this Manufacturing Agreement shall nonetheless continue
in full force and effect with the unenforceable provision(s) stricken
here from. In such case the applicable law shall apply with regard to
unenforceable and / or void provisions.
21.4 Governing Law. This Manufacturing Agreement shall be governed by and
construed in accordance with the laws of the United States and the
State of Delaware, excluding its choice of law rules. The Parties
consent to the jurisdiction of the state and federal courts of
Delaware.
22. Schedules. Should conflicts occur between this Manufacturing Agreement and
any part of the following Schedules, the requirements of this Manufacturing
Agreement shall govern. The following Schedules form an integral part of
this Manufacturing Agreement:
Schedule 1: Required Quality Data
Schedule 2: Portal Measurement Products and Specifications
Schedule 3: Philips Terms of Delivery
Schedule 4: Prices/Quantities
Schedule 5: Contact Persons
Schedule 6: Cosmetic Defect Guideline
Schedule 7: Cosmetic Inspection Guideline
13
DIAMETRICS MEDICAL, INC. PHILIPS MEDICAL SYSTEMS NORTH
AMERICA COMPANY, A DIVISION OF
PHILIPS ELECTRONICS NORTH
AMERICA CORPOPRATION
By: /s/ Xxxxx X. Xxxxxx By: /s/ Xxx Xxxxxxxx
----------------------------------- -----------------------------------
Xxxxx Xxxxxx, Chairman of the Board Xxx Xxxxxxxx
and Chief Executive Officer Vice-President & General
Manager - POCD
14
Schedule 1
Required Quality Data
Diametrics assures 5 years of record retention of all the following quality and
reliability related data:
Starting on the Effective Date, Diametrics shall provide the following data to
Philips Procurement once every quarter:
1. Outgoing data (Also known as F2 data) on each Portal Measurement Module:
100% outgoing inspection means
N = total (100%) each products tested.
2. YIELD (overall - from incoming through outgoing -
Philips expects a yield > 92%, per test station
@ manufacturing, 1/st/ cycle)
2.1) and a pareto analysis of all materials failures,1/st/ cycle.
2.2) and a pareto analysis of all process failures, 1/st/ cycle.
3. A pareto analysis of all root cause failures (sub-assembly level) for
returned products.
1
Schedule 2
Portal Measurement Products & Specifications
Item Part Number Respective Date of
Description Specification Specification
1. M3561A Portal Reader B-M3561-60000-1* 09/12/03
(M3561-6001)
2. M3561-68001 B-M3561-60000-1 09/12/03
Exchange repair Portal
3. M3565A temp card B-M3561-60000-1 09/12/03
(DMI 467900)
4. M3561-66403 edge connector
(DMI 516300) B-M3561-60000-1 09/12/03
5. M3564A DeviceSet w/DeviceCom
(DMI 464218) B-M3561-60000-1 09/12/03
6. M3672A edge connector
Cleaning kit (DMI 450000) N/A N/A
7. Cosmetic Defect Guideline see schedule 6
8. Cosmetic inspection see schedule 7
* B-M3561-60000-1 is the M3561A Portal Product Specification. Rev. A of such
specification applied prior to 9/12/03. To the extent of any conflict between
Rev. A or B of the M3561A Portal Product Specification and Diametrics'
manufacturing process document MP46000 Rev. X., MP46000 shall apply.
1
Schedule 3
Terms of Delivery
- Delivery and Packaging Standard of Philips
HP Dwg. A-5961-3658-1, Xxx. X
- XXX - capabilities for order receiving and acknowledgment
- Shipping date as specified on Philips-Purchase Order
- Date of dispatch: Philips Delivery Date minus 7 days
- Mode of transport : Airfreight
- Each shipment must indicate the exact address of the addressee on the outside
of the packaging.
- Products are shipped F.O.B. point of origin.
For Shipping and Production and Failure Analysis:
Philips Medizin Systeme Boeblingen Gmbh
Hewlett Xxxxxxx Xxx. 0
00000 Xxxxxxxxxx /Germany
POCD Production Engineer
Xxxxxxxx 0/ Xxxxx 0 0 X00
- Xxx minimum order and ship quantity is 10 pieces.
- Dangerous goods must be packed and labeled in accordance with the relevant
regulations (Postal regulations, Regulation on the transport of dangerous
goods, etc.).
- Each shipment must contain a delivery voucher indicating the Philips order
number and the Philips part
number.
- The total number of Portal Measurement Modules shipped shall be exactly as
specified on the Philips purchase order.
2
Schedule 4
Diametrics/Philips Transfer Prices
1. M3561A Portal Reader ***
(M3561-60001)
2. M3561-68000 Exchange repair Portal
The first two rates assume the unit is part of the exchange program, whereas the
third rate assumes the unit is not part of the exchange program.
- Combined repair/recondition rate (if outside warranty): ***
(assumes part of exchange program, and covers
reconditioning and functional repairs while
outside of the standard warranty period)
- Recondition only rate (would apply if within warranty): ***
(assumes part of exchange program, and covers
reconditioning and functional repairs while within
the standard warranty period, e.g., reconditioning
is not covered by warranty)
- Repair only rate (no recondition occurs): ***
(assumes not part of exchange program)
3. M3565A temp card ***
(DMI 467900)
4. M3561-66403 edge connector ***
(DMI 516300)
5. M3564A DeviceSet w/DeviceCom ***
(DMI 464218)
6. M3672A edge connector ***
Cleaning kit (DMI 450000)
7. M3561-40003 edge connector cover ***
(DMI 466400)
----------
*** Denotes confidential information that has been omitted from the
exhibit and filed separately, accompanied by a confidential treatment
request with the Securities and Exchange Commission pursuant to Rule
24b-2 of the Securities Exchange Act of 1934.
3
Schedule 5
Contact Persons
For Philips:
------------
Manufacturing e-mail: xxxxx.xxxxxxxxx@xxxxxxx.xxx
telephone:x00 (00000) 000-0000
FAX:x00(0000)000-0000
Purchasing Dept. e-mail: xxxxxxxx.xxxxxxxxx@xxxxxxx.xxx
telephone:x00(00000)000-0000
FAX:x00 (00000)000-0000
Materials Engineering e-mail: xxxxxx.xxxxxxxx@xxxxxxx.xxx
Telephone:x00(00000)000-0000
FAX:x00(0000)000-0000
Order Processing e-mail:xxxxxxxx.xxxxxxxxx@xxxxxxx.xxx
telephone:x00(00000)000-0000
FAX:x00(00000)000-0000
--------------------------------------------------------------------------------
For Diametrics:
---------------
Operations e-mail:xxxxxxx@xxxxxxxxxx.xxx
Xxxxx Xxxxxx telephone: 000-000-0000
FAX: 000-000-0000
Purchasing Dept. e-mail:xxxxxxxxx@xxxxxxxxxx.xxx
Xxxx Xxxxxxxx telephone: 000-000-0000
FAX: 000-000-0000
Materials Engineering/Manufacturing e-mail:xxxxxxxxxx@xxxxxxxxxx.xxx
Xxx Xxxxxxxxx telephone: 000-000-0000
FAX: 000-000-0000
Order Processing e-mail: xxxxxxxxx@xxxxxxxxxx.xxx
Xxx Xxxxxxxx telephone: 000-000-0000
FAX: 000-000-0000
4
Schedule 6
Cosmetic Defect Guideline
Exchange Portal Cosmetic Defect Guideline
Purpose
This document provides guidelines for inspecting used (also called exchange)
Portal Measurement Modules during the repair process. The overall purpose of
this guideline is to ensure that exchange products which are shipped to
customers are in a condition which will lead to consistent customer
satisfaction.
Scope
This document does not cover cosmetic specifications for individual recovered
parts of assemblies, nor cosmetic specifications for new whole units. This
guideline only applies to the Portal Measurement Module and its related exchange
parts.
Because this guide applies to refurbished product only, inspectors should be
examining the product for end-user and shipping induced damage, or wear, not
defects caused during manufacturing. Only cosmetic defects that could be caused
by end-user or shipment can fail the product since any manufacturing-induced
defects have already passed manufacturing inspection. For that reason,
manufacturing-caused cosmetic defects such as blush, flash, flow marks, haze,
sinks, and weld lines should not be considered as defects when examining
refurbished units.
General Guidelines
Cosmetics, in this context refer to superficial, visual appearance of the
product and as such do not affect the functioning of the product. Any defect
that interferes with the normal operation of the product, should be considered a
non-cosmetic defect. In the strictest sense, missing parts are not considered a
cosmetic defect, even if they are not needed for the operation of the product
(example: missing label). Note, however, that a missing part would be considered
a defect, and would necessitate some form of repair or action. Also, it is very
important to realize that cosmetics are largely subjective, so inspectors should
apply their best judgment in all cases.
The guiding principal when applying this judgment should be that this product is
not a new product and should not be considered as such. A balance needs to take
place between attempting to achieve near perfect cosmetic appearances which
would result in a large financial liability to the company versus the
possibility of having a high customer dissatisfaction rate if we were to ship
exchange product which looks shabby.
Evaluation Criteria
Cosmetic defects will be evaluated using a weighted criteria basis based on the
location of the defect and the severity of the defect. Guidance is provided in
Table 1.
Customer Added Labels
5
Any labels or marking put on a system by customers must be removed . This
includes such items as inventory tags and safety stickers. All glues used to
attach these labels must also be removed.
------------------------------------------------------------------------------------------------------------
Surface
Area
Location (a) Type of Cosmetic Defect
------------------------------------------------------------------------------------------
Small Deep Scratches
Paint Pits, Nicks Scratches (Very
Discoloration Heat Dis- Gouges- (minimally visible- Note
or wear Abrasion coloration (Note 2 visible) 2)
-------------------------------------------------------------------------------------------------------------
Product Minor OK- Minor OK- Minor OK- Minor OK Minor OK OK
Serial # Label
Note 1 Note 1 Note 1 Note 1 Note 1 Note 1
-------------------------------------------------------------------------------------------------------------
Logo "Portal" Very minor; Very minor; Very minor; None Very minor; None
(top)
Still Still clearly Still Still
clearly read name clearly clearly read
read name read name name
-------------------------------------------------------------------------------------------------------------
Display Screen
(glass area) N/A None N/A None None None
-------------------------------------------------------------------------------------------------------------
Case Top Minor OK Minor OK Minor OK None Minor OK None
-------------------------------------------------------------------------------------------------------------
Case Bottom Minor OK Minor OK Minor OK Minor OK Minor OK Minor OK
-------------------------------------------------------------------------------------------------------------
Note 1: Damage to the label that makes the reading of the serial number, product
number, impossible or difficult to read are not acceptable.
Note 2: Judgment needed here. A scratch, xxxxx, or xxxx, could be deep but small
in size (width and length). If this defect is small in size it may be acceptable
if in a location where seldom seen).
6
Schedule 7
Cosmetic Inspection Guideline
--------------------------------------------------------------------------------
Acceptance Criteria
--------------------------------------------------------------------------------
Checking Contents Packaging / box labeling
Portal
Edge connector cleaning kit
Certification sheet
Checking Portal Confirm Correct Label, Placement & Affixment for the
Unit
Ensure that there are 4 Rubber Feet and no abnormal
scuffing or damage, other than normal production
scuffs
Check for 4 torque screws on the bottom of Portal
Check for serial port cover and 2 torque screws
Check for light pipe
No scratches or fingerprints on screen
No dust between the touchscreen and display
No residues on portal
Sensor/IR probe flush or less
Discoloration, Flow marks, and Haze: Two per surface;
max dimension:0.5 mm (.02")
Weld Line Exceptions: Weld lines of up to 9.5mm
(.375") long shall be
allowed next to the IR Probe hole and the Light Pipe
hole in the M3561-40000 Top Housing
Pits and Specks: Two per surface; max
dimension: 0.5 mm (.02)
Scratches: Two per surface;
max dimension: 0.3 mm x 0.8 mm (.01" x .03")
Sinks: Two per surface; max dimension: 0.8mm (.03")
Weld Lines per surface: Two;
max dimension: 3 mm (.125")
No visual gasket btw. Touchscreen and top housing
Checking portal box Edge connector cleaning kit in Box
Certificate of Conformance
No crumb or dirt in the box
No bent connector pins
Confirm Correct Label, Placement & Affixment / for
the XXX
0