Conditions in the mist chamber. 5.1. The exposure zone of the mist chamber shall be maintained at 35 ± 5°C. At least two clean mist collectors shall be placed within the exposure zone, so that no drops of solution from the test samples or any other sources are collected. The collectors shall be placed near the test samples, one as near as possible to any nozzle and one as far as possible from all nozzles. The mist shall be such that, for each 80 cm2 of horizontal collecting area, from 1.0 to 2.0 ml of solution per hour, when measured over an average of at least 16 hours, is collected in each collector.

Appears in 15 contracts

Samples: wiki.unece.org, wiki.unece.org, wiki.unece.org

AutoNDA by SimpleDocs

Conditions in the mist chamber. 5.1. The exposure zone of the mist chamber shall be maintained at 35 ± + 5°C. ° C. At least two clean mist collectors shall be placed within the exposure zone, zone so that no drops of solution from the test samples or any other sources are collected. The collectors shall be placed near the test samples, one as near as possible nearest to any nozzle and one as far as possible furthest from all nozzles. The mist shall be such that, that for each 80 cm2 of horizontal collecting areaareas, there is collected in each collector from 1.0 to 2.0 ml of solution per hour, hour when measured over an average of at least 16 hours, is collected in each collector.

Appears in 1 contract

Samples: tind-customer-undl.s3.amazonaws.com

AutoNDA by SimpleDocs

Conditions in the mist chamber. 5.1. The exposure zone of the mist chamber shall be maintained at 35 ± 5°C5EC. At least two clean mist collectors shall be placed within the exposure zone, so that no drops of solution from the test samples or any other sources are collected. The collectors shall be placed near the test samples, one as near as possible to any nozzle and one as far as possible from all nozzles. The mist shall be such that, for each 80 cm2 of horizontal collecting area, from 1.0 to 2.0 ml of solution per hour, when measured over an average of at least 16 hours, is collected in each collector.

Appears in 1 contract

Samples: tind-customer-undl.s3.amazonaws.com

Draft better contracts in just 5 minutes Get the weekly Law Insider newsletter packed with expert videos, webinars, ebooks, and more!