In calibration definition

In calibration means making a reading with a calibrated test instrument within 6 inches of the site sensor. Verify that the sensor reading (via the permanent thermostat, gage or building automation system (BAS)) compared to the test instrument-measured value is within the tolerances specified in the prefunctional checklist requirements ( ). If not, install offset in BAS, calibrate or replace sensor. Use the same test instruments as used for the original calibration, if possible.
In calibration means observing a readout in the BAS and going to the actuator or controlled device and verifying that the BAS reading is correct. Fix items out of calibration or adjustment, via an offset in the BAS, or a mechanical fix.
In calibration means making a reading with a calibrated test instrument within 6 inches of the site sensor.

Examples of In calibration in a sentence

  • In calibration mode the display must indicate the maximal depression speed of the loading disc (v, mm/s) and the data to be checked determined by the manufacturer, for example: calibration factor and all values, which must be read out for the calibration.


More Definitions of In calibration

In calibration means observing a readout in the BAS and going to the actuator or controlled device and verifying that the BAS reading is correct.
In calibration means making a reading with a calibrated test instrument within 6 inches of the site sensor. Verify that the sensor reading (via the permanent thermostat, gage, or building automation system (BAS)) compared to the test instrument-measured value is within the tolerances specified in the Prefunctional Checklist requirements
In calibration means observing a readout in the BAS and going to the actuator or controlled device and verifying that the BAS reading is correct. For items out of calibration or adjustment, corrections may be applied in realtime during testing via an offset in the BAS or a mechanical fix if time permits, or be recorded in the Issues Log for follow-up by the construction team. Device or Actuator Procedure/ State BAS Value Instrument Measured/O bserved Value Pass Y/N Notes B-1 ISO Valve- Open Close - - - [1] B-2 ISO Valve- Open Close - - - [1] B-1 ON OFF On On Y B-2 ON OFF On On Y [2] B1&2 Relief Valves - - - - SAMPLE

Related to In calibration

  • Recalibration means the adjustment of all DRG weights to reflect changes in relative resource consumption.

  • Testing means that element of inspection that determines the properties or elements, including functional operation of materials, equipment, or their components, by the application of established scientific principles and procedures.