Calibration curves for MTT test Sample Clauses

Calibration curves for MTT test. Cell counting: Under sterile conditions 100 µl of cell suspension was mixed with an equal volume of Trypan blue dye by gentle pipetting. Trypan blue dye only stains dead cells. The number of viable cells was counted in a pre-cleaned haemocytometer and the concentration of viable cells was determined using the following equation: Concentration of viable cells (cells/ml) = A x B x C Where A is the mean number of viable cells counted, B is the dilution factor and C is the correction factor (104). Before performing an MTT test, calibration curves were constructed to determine the optimal cell seeding density for the experiment. The cells were seeded into a 96-xxxxx plate at concentrations of 5x105, 2.5x105, 1.25x105, 6.25x104, 3.125x103, 1.56x103, 7.81x102, 3.91x102 and 1.95x 101 cells/ml. The experiment blank consisted of 100 µl of medium only. The cells were incubated for 24hrs at 37 ºC, 5% CO2. Solutions: A 1 mg/ml MTT solution was made by dissolving 40 mg of MTT powder in 40 ml sterile PBS (the mixture was sonicated for a few seconds to aid dissolution) and stored in the fridge (2-8 C). SDS solution was prepared by dissolving 4g of SDS powder in 20 ml pure water. After sonication, 20ml of DMF was added to the SDS solution to give a final concentration of 10% SDS. After 24 hr incubation, 25 µl of filtered MTT solution (1mg/ml) mixed with 25µl medium was added to each well and the cells re-incubated for 4hrs. Normally the dissolved MTT is converted to insoluble purple formazan crystals by cleavage of the tetrazolium ring by dehydrogenase enzymes in viable cells. This water insoluble formazan is solubilized using SDS. At the end of the incubation period, the medium was removed from the plate and the converted dye was solubilized by the addition of 100 µl of SDS solution to each well (including the blank). The cultures were incubated overnight and absorbance was measured at two wavelengths; 570 and 650 nm. A calibration curve was constructed by plotting number of cells against the mean absorbance (Appendix).
AutoNDA by SimpleDocs

Related to Calibration curves for MTT test

  • Calibration The comparison of a measurement system or device of unverified accuracy with a measurement system of known and greater accuracy to detect deviation of the unverified measurement system from required performance specifications (of the unverified measurement system or device) and to quantify all measured values to applicable units of the international system of units.

  • Meter Testing Company shall provide at least twenty-four (24) hours' notice to Seller prior to any test it may perform on the revenue meters or metering equipment. Seller shall have the right to have a representative present during each such test. Seller may request, and Company shall perform, if requested, tests in addition to the every fifth-year test and Seller shall pay the cost of such tests. Company may, in its sole discretion, perform tests in addition to the fifth year test and Company shall pay the cost of such tests. If any of the revenue meters or metering equipment is found to be inaccurate at any time, as determined by testing in accordance with this Section 10.2 (Meter Testing), Company shall promptly cause such equipment to be made accurate, and the period of inaccuracy, as well as an estimate for correct meter readings, shall be determined in accordance with Section 10.3 (Corrections).

  • Performance Testing 7.2.1 The Design-Builder shall direct and supervise the tests and, if necessary, the retests of the Plant using Design-Builder’s supervisory personnel and the Air Emissions Tester shall conduct the air emissions test, in each case, in accordance with the testing procedures set forth in Exhibit A (the “Performance Tests”), to demonstrate, at a minimum, compliance with the Performance Guarantee Criteria. Owner is responsible for obtaining Air Emissions Tester and for ensuring Air Emissions Tester’s timely performance. Design-Builder shall cooperate with the Air Emissions Tester to facilitate performance of all air emissions tests. Design-Builder shall not be held responsible for the actions of Owner’s employees and third parties involved in the Performance Testing, including but not limited to Air Emissions Tester.

  • Temperature Measurement Temperature will be measured by the nearest automatic Melbourne Bureau of Meteorology Monitoring Station for example (but not limited to): Melbourne, Moorabbin, Dunns Hill, Melbourne Airport, Frankston, and Point Xxxxxx. At the commencement of each project, the onsite management and employee representatives shall agree which is to be the applicable automatic weather monitoring station.

  • Outputs 11. The objectives and outcomes of this Agreement will be achieved by:

  • Completion of Evaluation Cycle 1. The summative evaluation rating shall be based upon a preponderance of the evidence, assessed in a holistic manner, that is aligned to the Ohio Educator Standards. Only evidence gathered during the walkthroughs and formal observations that are conducted for the current school year may be used.

  • Test The outer surface of the lamp lenses shall be subjected once or more than once to the action of the sand jet produced as described above. The jet shall be sprayed almost perpendicular to the surface to be tested. The deterioration shall be checked by means of one or more samples of glass placed as a reference near the lenses to be tested. The mixture shall be sprayed until the variation in the diffusion of light on the sample or samples measured by the method described in Appendix 2, is such that: Δ d = (T5 - T4) / T2 ≤ 0.0250 ± 0.0025 Several reference samples may be used to check that the whole surface to be tested has deteriorated homogeneously. Annex 6 - Appendix 4

  • Repair Testing At the time of repair of a LIS trunk group, at no additional charge, tests will be performed to ensure that the service is operational and meets the applicable technical parameters.

  • Test Samples The Contractor is responsible for providing Samples of sufficient size for test purposes and for coordinating such tests with the Work Progress Schedule to avoid delay.

  • Interconnection Customer Drawings Within one hundred twenty (120) days after the date of Initial Operation, unless the Interconnection Parties agree on another mutually acceptable deadline, the Interconnection Customer shall deliver to the Transmission Provider and the Interconnected Transmission Owner final, “as-built” drawings, information and documents regarding the Customer Interconnection Facilities, including, as and to the extent applicable: a one-line diagram, a site plan showing the Customer Facility and the Customer Interconnection Facilities, plan and elevation drawings showing the layout of the Customer Interconnection Facilities, a relay functional diagram, relaying AC and DC schematic wiring diagrams and relay settings for all facilities associated with the Interconnection Customer's step-up transformers, the facilities connecting the Customer Facility to the step-up transformers and the Customer Interconnection Facilities, and the impedances (determined by factory tests) for the associated step-up transformers and the Customer Facility. As applicable, the Interconnection Customer shall provide Transmission Provider and the Interconnected Transmission Owner specifications for the excitation system, automatic voltage regulator, Customer Facility control and protection settings, transformer tap settings, and communications.

Time is Money Join Law Insider Premium to draft better contracts faster.