Temperature Calibration Clause Samples

Temperature Calibration. Refer to the Service Mode section of this manual for a detailed explanation of temperature calibration. For Health Care sterilizers use the Steam calibration method. For Scientific units use the Oil Bath calibra- tion method, where required; otherwise the Steam method may be used.
Temperature Calibration. The best standard for TGA temperature calibration is the use of ferromagnetic standards. These materials undergo a reversible, second-order phase transition from the ferromagnetic to the paramagnetic state when a critical temperature (▇▇▇▇▇ temperature) is exceeded and loose the strong magnetization typical of ferromagnetic materials. A magnet appropriately placed above or below the furnace provides the magnetic field that attracts the reference materials placed in the sample holder. When going through the ▇▇▇▇▇ transition the reference material magnetization vanishes resulting in an apparent weight change. Each ferromagnetic material undergoes the ferromagnetic-paramagnetic transition at a reproducible, characteristic temperature that is only determined by the chemical composition of the ferromagnetic sample. A working group under the auspices of the International Congress for Analysis and Calorimetry (ICTAC) selected a series of alloys having ▇▇▇▇▇ temperatures approximately every 200 ºC between 152 and 1116 ºC and the ▇▇▇▇▇ temperatures of these reference materials were determined in an interlaboratory test (▇▇▇▇▇▇▇▇▇, 2003). These materials (alloys of cobalt and nickel) are available from TA Instruments. Alternatively, ▇▇▇▇▇ Point calibration kits spanning between 163°C (Alumel alloy) and 1000°C (Hisat-50 alloy) are available from ▇▇▇▇▇▇-▇▇▇▇▇.
Temperature Calibration. 2.4.1 A temperature bath is prepared at a specified range. Using a CTD to monitor the temperature the ▇▇▇▇▇▇ is allowed to soak while the heater/chiller is allowed to run. The computer polls both the CTD and ▇▇▇▇▇▇ for the values of their onboard thermometers, these values are then stored for processing. After running the desired temperature range the data is plotted and a linear relationship between the two data sets is found. The resultant gain and offset values are then used as coefficients for ▇▇▇▇▇▇’▇ temperature sensor.

Related to Temperature Calibration

  • Calibration The comparison of a measurement system or device of unverified accuracy with a measurement system of known and greater accuracy to detect deviation of the unverified measurement system from required performance specifications (of the unverified measurement system or device) and to quantify all measured values to applicable units of the international system of units.

  • Temperature Where low temperature and/or self-service cases are used for any of such merchandise coming under the jurisdiction of the Union, such cases shall be served only by employees covered by this Agreement.

  • Temperature Measurement Temperature will be measured by the nearest automatic Melbourne Bureau of Meteorology Monitoring Station for example (but not limited to): Melbourne, Moorabbin, Dunns Hill, Melbourne Airport, Frankston, and Point ▇▇▇▇▇▇. At the commencement of each project, the onsite management and employee representatives shall agree which is to be the applicable automatic weather monitoring station.

  • Meters 7.1.1 For installation of Meters, Meter testing, Meter calibration and Meter reading and all matters incidental thereto, the SPD and SECI shall follow and be bound by the Central Electricity Authority (Installation and Operation of Meters) Regulations, 2006, the Grid Code, as amended and revised from time to time. 7.1.2 The SPD shall bear all costs pertaining to installation, testing, calibration, maintenance, renewal and repair of meters at SPD’s side of Delivery Point. 7.1.3 In addition to ensuring compliance of the applicable codes, the SPD shall install Main & Check meters at the Delivery Point, along with Stand-by meter(s) as per the applicable Central/State regulations. 7.1.4 In case of pooling of multiple Projects, power from multiple Projects can be pooled at a Pooling Substation prior to the Delivery point and the combined power can be fed at Delivery point through a common transmission line from the Pooling Substation. In such cases, ABT compliant sub-meters as per relevant regulation/approval are also to be set up at pooling substation for individual projects in addition to the meters at Delivery Point as described in clause 7.1.3.

  • Outputs Analogue and digital outputs of protected content are allowed if they meet the requirements in this section and if they are not forbidden elsewhere in this Agreement..