Laboratory Quality Control Samples Sample Clauses

Laboratory Quality Control Samples. Laboratories routinely perform medium spike and laboratory duplicate analysis on field samples as a quality control check. A minimum of one field sample per week or 1 per 20 samples (including field blanks and duplicates), whichever is greater, must be designated as the "Lab QC Sample" for the medium and laboratory duplicate analysis. Laboratory quality control samples should be selected from sampling points which are suspected to be moderately contaminated. Label the bottles and all copies of the paperwork as "Lab QC Sample"; the laboratory must know that this sample is for their QC analyses. The first laboratory QC sample of the sampling effort should be part of the first or second day's shipment. Subsequent laboratory QC samples should be spread out over the entire sampling effort. For water media, 2-3 times the normal sample volume must be collected for the laboratory QC sample. Additional volume is usually not necessary for soil samples.
AutoNDA by SimpleDocs
Laboratory Quality Control Samples. For each batch of samples, additional samples are prepared at the laboratory. These samples include: method blanks, low concentration or control check standards, laboratory check standards, matrix spike and matrix spike duplicate (MS/MSD) results. Blank or low concentration spike samples and duplicates are prepared to determine if any contamination has occurred in the laboratory and to determine the percent recovery at low concentrations. Preservative blanks are also prepared in the laboratory to determine if any contamination was introduced from the preservatives. The MS/MSD samples are used to determine percent recoveries of the reference standards and matrix spikes, and are used to detect matrix interferences. The laboratory duplicate samples are used to determine the relative percent differences, which can be used to detect laboratory equipment problems such as drift in calibration. The frequency of these QA/QC samples and actions that can be taken are shown in Table 5 for mercury and in Table 6 for the other chemical ancillary parameters. A quality assurance/quality control (QA/QC) summary form will be completed by the laboratory for water and sediment samples. The sample numbers for all QA/QC rinseate samples, reference samples, laboratory QC samples, and duplicates will be documented on this form so that the QA/QC samples can be associated with each batch of environmental samples. QA Sample QA Measure Quality Control Criteria for Analysis of Mercury Minimum Frequency Acceptance Limits Corrective Action* Total and Dissolved Iron Method Blank Limit Study Method Blank Spiked at PQL Determine MDL Determine PQL Run 20 Method Blanks per batch of samples Run 20 spiked samples per batch of samples MDL is 3X standard NA deviation of mean of 20 blanks PQL is 4 X MDL or mean NA of spiked blanks Method Blanks Analyzed with Samples Accuracy 3 per analytical batch, whichever contains fewer samples Analyte >MDL Check reagents, clean instrument, prepare fresh standards and reanalyze samples Certified Reference Material (CRM) Accuracy Once per analytical batch Analyte Specific 75%< CRM<125% THg; 70%< CRM <130% MeHg Check reagents and clean instrument, and reanalyze samples. Accepting the data and acknowledging the level of uncertainty with a written explanation Replicate Precision Once every 10 samples or every analytical batch, whichever contains fewer samples Analyte Specific 75%< %D <125% THg; 70%< %D <130% MeHg Reanalysis of samples Accepting the data and acknow...
Laboratory Quality Control Samples. A minimum of one field sample per week or 1 per 20 samples (including field blanks and duplicates), whichever is greater, must be designated as the "Lab QC Sample" for the matrix and laboratory duplicate analysis. Laboratory quality control samples should be selected from sampling points which are suspected to be moderately contaminated. Label the bottles and all copies of the paperwork as "Lab QC Sample"; the laboratory must know that this sample is for their QC analyses. The first laboratory QC sample of the sampling effort should be part of the first or second day's shipment. Subsequent laboratory QC samples should be spread out over the entire sampling effort. For water matrices, 2-3 times the normal sample volume must be collected for the laboratory QC sample.

Related to Laboratory Quality Control Samples

  • Quality Assurance/Quality Control Contractor shall establish and maintain a quality assurance/quality control program which shall include procedures for continuous control of all construction and comprehensive inspection and testing of all items of Work, including any Work performed by Subcontractors, so as to ensure complete conformance to the Contract with respect to materials, workmanship, construction, finish, functional performance, and identification. The program established by Contractor shall comply with any quality assurance/quality control requirements incorporated in the Contract.

  • Quality Control A. Controlled Affiliate agrees to use the Licensed Marks and Name only in connection with the licensed services and further agrees to be bound by the conditions regarding quality control shown in attached Exhibit A as they may be amended by BCBSA from time-to-time.

  • Quality control system (i) The Contractor shall establish a quality control mechanism to ensure compliance with the provisions of this Agreement (the “Quality Assurance Plan” or “QAP”).

  • Use; Quality Control a. Neither party may alter the other party’s trademarks from the form provided and must comply with removal requests as to specific uses of its trademarks or logos.

  • Quality Control Program The Contractor shall describe the Quality Control Program in a written document which shall be reviewed by the Engineer prior to the start of any production, construction, or off-site fabrication. The written Quality Control Program shall be submitted to the Engineer for review at least ten (10) calendar days before the start of construction. The Quality Control Program shall be organized to address, as a minimum, the following items:

  • Quality Control Plan Within forty five (45) days after the design development documents are approved, the Design-Builder shall develop a quality control plan for the Project (the “Quality Control Plan”). A draft of the Quality Control Plan shall be submitted to the Department and shall be subject to the Department’s review and approval. The Quality Control Plan shall be tailored to the specific products/type of construction activities contemplated by the GMP Basis Documents, and in general, shall include a table of contents, quality control team organization, duties/responsibilities of quality control personnel, submittal procedures, inspection procedures, deficiency correction procedures, documentation process, and a list of any other specific actions or procedures that will be required for key elements of the Work.

  • DEMONSTRATIONS/SAMPLES 9.1 Bidders shall demonstrate the exact item(s) proposed within seven (7) calendar days from receipt of such request from the Owners.

  • Data Quality 4.1 Each party ensures that the shared Personal Data is accurate.

  • Erosion and Sediment Control The Contractor shall install and maintain erosion and sediment control devices to prevent adjacent streets, storm drains and property from accumulations of soil, sediment, or dust which result from his work. Devices installed to control sediment and prevent erosion shall comply with the requirements of the Georgia Tech Project Superintendent. The Contractor is responsible for monitoring downstream conditions throughout the construction period and clearing any debris, sediment, and dust caused by the progress of the Work. The Contractor shall inspect, maintain, and clear erosion and sediment control devices at least weekly unless otherwise directed by the Georgia Institute of Technology Construction Manager. At the conclusion of the Work, the Contractor shall remove all installed devices and restoring the site to the satisfaction of the Georgia Tech Construction Manager.

  • Supervisory Control and Data Acquisition (SCADA) Capability The wind plant shall provide SCADA capability to transmit data and receive instructions from the ISO and/or the Connecting Transmission Owner for the Transmission District to which the wind generating plant will be interconnected, as applicable, to protect system reliability. The Connecting Transmission Owner for the Transmission District to which the wind generating plant will be interconnected and the wind plant Developer shall determine what SCADA information is essential for the proposed wind plant, taking into account the size of the plant and its characteristics, location, and importance in maintaining generation resource adequacy and transmission system reliability in its area.

Draft better contracts in just 5 minutes Get the weekly Law Insider newsletter packed with expert videos, webinars, ebooks, and more!